Write a Review

Radwag
Radwag WL-307-0006 PMV 50.5Y Moisture Analyzer
Radwag WL-307-0006 PMV 50.5Y Moisture Analyzer
- Brand
- : Radwag
- SKU
- : WL-307-0006
- UPC
- :
- Condition
- : New
Radwag WL-307-0006 PMV 50.5Y Moisture Analyzer
- microwave moisture analyzer can be successfully used wherever materials containing large amounts of water are dried. These materials are, for example, dairy products, fish, meat, fruit products, syrups, creams or liquid resin.
- The sample is placed in the drying chamber on a glass fibre filter. At the bottom of the chamber there is one end of the waveguide from where the microwaves are emitted. Upon process start, microwaves are equally distributed on the filter’s surface and dry the sample. The moisture content is determined on the basis of the mass loss obtained during drying.
- A respective drying program is selected depending on sample mass and its chemical composition. The average duration of this process depends on the type and mass of used substance, and takes 1 to 10 minutes. In contrast, drying using the traditional halogen moisture analyzer takes 5 to 40 minutes.
- The PMV 50.5Y microwave moisture analyzer is equipped with a temperature sensor and enables to monitor device power in order to prevent exceeding the boiling point during the process. A preset microwave power is displayed on the indicator throughout the drying procedure.
- intended for drying of samples containing significant amount of moisture (up to 100%),
- ultra-short drying time when compared to traditional moisture analyzers and standard drying methods,
- microwave-based solution,
- uniform heating of entire sample’s surface,
- complex database of drying programs,
- finish mode: manual, automatic, time-defined, user-defined,
- advanced reports, time and statistical graphs,
- easy operation due to a touch screen.
- 60 glass fibre filters of ø90 mm,
- a sample of sodium chloride 10% ± 0,5% to determine correctness of the operation at the workstation,
- a service concerning determination of drying parameters for five customer-selected samples carried out by RADWAG employee.
The results obtained using the moisture analyzer (the final result of the drying process):
% M – the mass loss in %; displays the mass change recorded during the drying process, expressed as a percentage,
% D – the dry mass obtained during the drying process, expressed as a percentage; the result is the portion of the sample that remains on the weighing pan after the moisture has evaporated,
% R – the ratio of moisture content to dry mass, obtained during the drying process, expressed as a percentage; the result is the portion of the sample that evaporated during the drying process,
g – the sample mass after drying.
Terminal 5Y in Microwave Moisture Analyzer
- 10-inch display,
- Digital Weighing Auditor (DWA),
- uncompromised user verification,
- Ambient Light,
- hotspot,
- RFID*,
- Live Note text notes,
- voice commands for the device,
- measurement history within the Handy Library,
- widgets.
Modbus TCP/IP
- is one of the most popular communication protocols in industrial automation. Implementing it in the 5Y balances and CY10 scales makes the devices of this series friendly to automation integrators.
- A balance or scale in a Modbus system is a slave device and performs tasks specified by the master device which is most often a PLC. Protocol Modbus TCP/IP in the 5Y balances and C10 scales enables, among other things, the realization of mass readout, zeroing and tare functions of the device, as well as tare settings and MIN and MAX thresholds.
Product Specifications:
Metrological Parameters Maximum Capacity [Max] : 50 g Physical Parameters Leveling System : Semi-Automatic – LevelSENSING Construction Protection Class : IP 43 |
Communication Interface Communication Interface : 2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot Electrical Parameters Power Supply : 100 – 120 V AC 50/60 Hz or 200-240 V AC 50/60 Hz Environmental Conditions Operating Temperature : +10 – +40 °C SOP – Metrological Tests (additional options) Mass Indication Control : External mass standard 50 g, 100 g, 200 g, depending on the type of moisture analyzer (additional equipment) Features of Use Finish Mode : 4 modes (automatic, manual, time-defined, user-defined) |