Metrological Parameters
Maximum Capacity [Max] : 120 / 250 g Minimum Load : 1 mg Readability [d] : 0.01 / 0.1 mg Verification Unit [e] : 1 mg Tare Range : -250 g Standard Repeatability [5% Max] : 0.005 mg Standard Repeatability [Max] : 0.06 mg Standard Minimum Weight (USP) : 10 mg Standard Minimum Weight (U=1%, k=2) : 1 mg Permissible Repeatability [5% Max] : 0.012 mg Permissible Repeatability [Max] : 0.1 mg Linearity : ±0.06 / 0.2 mg Eccentric Load Deviation : 0.2 mg Sensitivity Time Drift : 1×10⁻⁶/Year×Rt Stabilization Time : 3 s Adjustment : Internal (Automatic) OIML Class : I
Physical Parameters
Leveling System : Automatic – Reflex Level System Display : 10″ Graphic Colour Touchscreen Weighing Chamber : Automatic Weighing Chamber Doors : Automatic Delivery Components : Analytical Balance, Weighing Pan, Weighing Pan Shield, Centring Ring, Brush, Fabric Dust Cover, Power Supply. Weighing Chamber Dimensions : 200×170×220 mm Weighing Pan Dimensions : ø90 open-work pan + ø85 (option) mm Packaging Dimensions : 750×492×595 mm Net Weight : 14.7 kg Gross Weight : 19.1 kg
Construction
Protection Class : IP 43
|
Communication Interface
Communication Interface : 2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot
Electrical Parameters
Power Supply : Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A Balance: 12 – 15V DC 1.6A max*
Environmental Conditions
Operating Temperature : +10 – +40 °C
Operating Temperature Change Rate : ±0.3 °C / 1 h (±1 °C / 8 h)
Relative Humidity : 20% – 80%.
Relative Humidity Change Rate : ±1% / h (±4% / 8 h)
|